inforMED
MalfunctionDCK

C-REACTIVE PROTEIN (LATEX)

Received Aug 2, 2001 · Event occurred Aug 1, 2001

Report 1823260-2001-00166 · MDR key 347104

Device

Generic name

Reagent Cassette

Manufacturer

Roche Diagnostics

Model number

NA

Catalog number

0764930

Lot number

156081-01

Product problems

  • Design/structure problem
  • Nonstandard Device
  • Incorrect, Inadequate or Imprecise Result or Readings

Patient

UNKNOWN

  • Unknown (for use when the patient's condition is not known)

Narrative

Description of Event or Problem

NOTIFIED BY GLOBAL AFFILIATE THAT REAGENT CASSETTES MAY CONTAIN UNDER-FILLED OR NEARLY EMPTY REAGENT BOTTLES. ANALYZER DOES NOT CONTAIN LEVEL DETECTION SYSTEM THEREFORE NO ALARM WOULD WARN CUSTOMERS. PATIENT SAMPLES AND CONTROLS MAY PRODUCE FALSELY LOW OR HIGH RESULTS WITHOUT AN ALARM OR THERE COULD BE UNACCEPTABLE PRECISION WHEN PATIENT OR CONTROL MATERIAL IS REPORTED.

Remedial action

  • Recall