MalfunctionLYK

2.3MM 1.0MM 80CM DISPOSABLE ANGIOSCOPE

Received Oct 11, 2013 · Event occurred Aug 9, 2013

Report 2027111-2013-00383 · MDR key 3466447

Device

Generic name

None

Manufacturer

Applied Medical

Model number

ANG-080-D10K

Catalog number

100326502

Lot number

1174936

Product problems

  • Unable to Obtain Readings

Patient

Not reported

  • No Known Impact Or Consequence To Patient

Narrative

Description of Event or Problem

LEFT FEM-POP BYPASS WITH INSERTION OF SAPHENOUS VEIN: "DISPOSABLE ANGIOSCOPE WOULD NOT DISPLAY PICTURE DURING SURGICAL PROCEDURE."

Additional Manufacturer Narrative

INVESTIGATION SUMMARY: THE EVENT UNIT WAS RETURNED FOR EVAL. UPON INSPECTION, ENGINEERING CONFIRMED THE DEVICE DISPLAYED A SKEWED AND BLURRY IMAGE. NO DAMAGE WAS NOTED TO THE UNIT. ENGINEERING DISASSEMBLED THE UNIT AND FOUND THAT ONE EYEPIECE LENS WAS OUT OF ITS PROPER POSITION. ALL ENDOSCOPES UNDERGO 100% VISUAL INSPECTION AND ARE FUNCTIONALLY TESTED DURING THE MANUFACTURING PROCESS. IT IS LIKELY THAT THE LENS WAS NOT SECURED IN ITS PROPER POSITION DUE TO AN OPERATOR ERROR AND BECAME DETACHED INSIDE THE EYEPIECE DURING SHIPPING AND HANDLING OF THE PRODUCT. THE PRODUCTION TEAM HAS BEEN NOTIFIED OF THIS OCCURRENCE IN ORDER TO DEVELOP A HEIGHTENED AWARENESS OF THE INCIDENT. THIS DOCUMENT REPRESENTS OUR INITIAL/FINAL REPORT.