2.3MM 1.0MM 80CM DISPOSABLE ANGIOSCOPE
Received Oct 11, 2013 · Event occurred Aug 9, 2013
Report 2027111-2013-00383 · MDR key 3466447
Device
Generic name
None
Manufacturer
Applied MedicalModel number
ANG-080-D10KCatalog number
100326502
Lot number
1174936
Product problems
- Unable to Obtain Readings
Patient
Not reported
- No Known Impact Or Consequence To Patient
Narrative
Description of Event or Problem
LEFT FEM-POP BYPASS WITH INSERTION OF SAPHENOUS VEIN: "DISPOSABLE ANGIOSCOPE WOULD NOT DISPLAY PICTURE DURING SURGICAL PROCEDURE."
Additional Manufacturer Narrative
INVESTIGATION SUMMARY: THE EVENT UNIT WAS RETURNED FOR EVAL. UPON INSPECTION, ENGINEERING CONFIRMED THE DEVICE DISPLAYED A SKEWED AND BLURRY IMAGE. NO DAMAGE WAS NOTED TO THE UNIT. ENGINEERING DISASSEMBLED THE UNIT AND FOUND THAT ONE EYEPIECE LENS WAS OUT OF ITS PROPER POSITION. ALL ENDOSCOPES UNDERGO 100% VISUAL INSPECTION AND ARE FUNCTIONALLY TESTED DURING THE MANUFACTURING PROCESS. IT IS LIKELY THAT THE LENS WAS NOT SECURED IN ITS PROPER POSITION DUE TO AN OPERATOR ERROR AND BECAME DETACHED INSIDE THE EYEPIECE DURING SHIPPING AND HANDLING OF THE PRODUCT. THE PRODUCTION TEAM HAS BEEN NOTIFIED OF THIS OCCURRENCE IN ORDER TO DEVELOP A HEIGHTENED AWARENESS OF THE INCIDENT. THIS DOCUMENT REPRESENTS OUR INITIAL/FINAL REPORT.