InjuryCAP
AIRWAY PRESSURE MONITOR
Received Jul 18, 1996 · Event occurred Jun 7, 1996
Report 1317346-1996-00002 · MDR key 34441
Device
Generic name
Airway Pressure Monitor
Manufacturer
Canadian Monaghan LtdModel number
CM5000SCCatalog number
900-50SC-001
Lot number
P620N
Product problems
- Device Alarm System
Patient
65 YR
- Apnea
Narrative
Description of Event or Problem
CUSTOMER REPORTS AIRWAY PRESSURE MONITOR DID NOT ALARM DURING A PT DISCONNECT. DELAY SETTING WAS AT 20 SECS. PT WAS DISCOVERED IN A HYPOXIC, CYANOTIC CONDITION BEFORE DISCONNECT WAS FOUND. PT CIRCUIT WAS RECONNECTED AND AIRWAY PRESSURE MONITOR REPLACED. SYSTEM THEN CHECKED OKAY. NO FURTHER PROPLEMS NOTED.