inforMED
MalfunctionILM

PARKER BATH

Received Sep 23, 2013 · Event occurred Dec 10, 2012

Report 1419652-2013-00268 · MDR key 3420692

Device

Model number

AL14110-CA

Product problems

  • Temperature Problem

Patient

Not reported

  • No Patient Involvement

Narrative

Description of Event or Problem

IT WAS REPORTED THAT; "TEMPERATURE FLUCTUATING AND TOO HIGH, 50+ DEGREES C. SPOKE WITH HEAD MAINTENANCE, MAINTENANCE PERSONNEL DID NOT KNOW HOW TO TAKE OUT AND CLEAN THE CARTRIDGE. MAINTENANCE TURNING UP THE TEMPERATURE BECAUSE IT WAS TOO LOW, TEMPERATURE WAS TOO LOW BECAUSE THE CARTRIDGES WERE DIRTY. CLEAN FILL CARTRIDGE AND ADJUSTED THE FILL TEMPERATURE TO 42 DEGREES C. EXPLAINED TO HEAD MAINTENANCE TO EDUCATE MAINTENANCE STAFF NOT TO ADJUST THE TEMPERATURE WITHOUT CLEANING THE FILL CARTRIDGE." REFERENCE MFT REPORT # (B)(4).