inforMED

SAFHS SONIC ACCELERATED FRACTURE HEALING SYSTEM

Received Feb 28, 1995

Report MW5000020 · MDR key 34175

Device

Generic name

Fracture Healing System

Manufacturer

Exogen, Inc.

Product problems

  • No Device Output
  • Material Separation
  • Transducer failure

Patient

NO INFO

  • Unknown (for use when the patient's condition is not known)

Narrative

Description of Event or Problem

A DEVICE FAILURE WAS REPORTED ON 2/10/95. A REPLACEMENT UNIT WAS SENT TO THE PT ON THE SAME DAY IN A PRIORITY OVERNIGHT SHIPMENT. UPON RETURN OF THE FAILED UNIT, VISUAL INSPECTION REVEALED THAT THE TRANSDUCER ELEMENT HAD SEPARATED FROM THE TRANSDUCER CUP. THIS CAUSED ONE OF THE TRANSDUCER INTERNAL CONNECTIONS TO SEPARATE. THIS PROBLEM CAUSED THE MAIN OPERATING UNIT (MOU) TO STOP OPERATING AND ACTIVATED THE TROUBLE ALARM AS THE SYSTEM WAS DESIGNED TO DO. THE TRANSDUCER HAS BEEN RETURNED TO THE MFR FOR EVALUATION OF THIS PROBLEM. THIS IS AN ISOLATED INCIDENT AND NO ACTION WILL BE INSTITUTED AT THE PRESENT TIME.