MalfunctionGDF

INSTRUMENT MCL55 PISTOL W/NEEDLE MCL55-1 [MXI/XOM]

Received Aug 28, 2013 · Event occurred Jun 27, 2012

Report 9680837-2013-00466 · MDR key 3314899

Device

Generic name

Guide, Needle, Surgical

Model number

MCL55

Catalog number

MCL55

Lot number

200004MF

Product problems

  • Break

Patient

Not reported

  • No Known Impact Or Consequence To Patient

Narrative

Description of Event or Problem

IT WAS REPORTED THAT AN MCL55 SLIPS UNDER PRESSURE. NO PATIENT IMPACT WAS REPORTED.

Additional Manufacturer Narrative

(B)(4). RESULTS IS MECHANICAL SHOCK PROBLEM. THE RETURNED DEVICE WAS EVALUATED AND CONFIRMED TO BE OUT OF SPECIFICATION. FUNCTIONAL TESTING WAS PERFORMED; MISSING AND DEFECTIVE COMPONENTS WERE REPLACED. NOTE: THIS MDR IS BEING FILED AS A RESULT OF AN INTERNAL REVIEW OF COMPLAINTS FROM MEDTRONIC¿S MXI FACILITY IN (B)(4). THIS REVIEW WAS CONDUCTED TO RESOLVE SEVERAL ISSUES DISCOVERED IN THE MXI COMPLAINT HANDLING PROCESS. ISSUES RESOLVED INCLUDE THE MISCLASSIFICATION OF DEVICES RETURNED FOR REPAIR, AS WELL AS GAPS IN REPORTING. MXI CAPA (B)(4) WAS OPENED TO ADDRESS THESE ISSUES.