inforMED
InjuryKNM

FLOWTRON EXCEL / TRIO PUMPS

Received Aug 15, 2013 · Event occurred Jul 16, 2013

Report 1419652-2013-00215 · MDR key 3307756

Device

Model number

247001

Product problems

  • Arcing

Patient

Not reported

  • Electric Shock

Narrative

Description of Event or Problem

CAREGIVER WAS SHOCKED WHILE ATTEMPTING TO PLUG IN AN AC550 IN A DIMLY LIT ROOM. WHAT HAD HAPPENED PREVIOUSLY IS THAT THE WARNING TAG ON THE PLUG WAS TORN OFF, BUT THE METAL EYELET REMAINED. THE EYELET SOMEHOW CONTRIBUTED TO THE SHOCK, GETTING INTO CONTACT WITH THE GROUND, ETC. FACILITY REP STATES THAT THEY BELIEVE THE DEVICE IS FUNCTIONING NORMALLY. AC500 PLUG ARCED AS CAREGIVER PLUGGED PUMP INTO APPROVED HOSPITAL GRADE GROUNDED ELECTRICAL OUTLET. CAREGIVER RECEIVED ELECTRICAL SHOCK, WAS EVALUATED AT ED, AN CLEARED WITH NO INJURIES. REF # 3005619970-2013-00009.