EO GAS STERILIZER
Received Aug 9, 2013 · Event occurred Jul 10, 2013
Report 3005899764-2013-00081 · MDR key 3274828
Device
Generic name
Sterilizer
Manufacturer
Steris Mexico, S. De R.l. De C.v.Product problems
- Chemical Spillage
- Gas/Air Leak
Patient
Not reported
- No Known Impact Or Consequence To Patient
Narrative
Description of Event or Problem
THE USER FACILITY REPORTED AN ETO LEAK. NO INJURIES TO HOSPITAL STAFF OR PATIENTS WERE REPORTED. THE DEPARTMENT WAS EVACUATED AND THE FIRE DEPARTMENT WAS DISPATCHED. NO PROCEDURAL DELAYS OR CANCELLATIONS OCCURRED.
Additional Manufacturer Narrative
THE USER FACILITY'S 3RD PARTY TECHNICIAN DETACHED THE EXHAUST MANIFOLD ASSEMBLY FROM THE EO STERILIZER WHILE REPAIRING THE STERILIZER. THE EXHAUST MANIFOLD IS SHARED WITH A SECOND EO STERILIZER, AND WHEN THE SECOND STERILIZER WAS OPERATED, EO GAS LEAKED FROM THE FIRST STERILIZER DUE TO DISCONNECTION OF THE EXHAUST MANIFOLD ASSEMBLY FROM THE FIRST STERILIZER. THE USER FACILITY REQUESTED A STERIS TECHNICIAN TO INSPECT AND ASSESS THE INCIDENT; THE STERIS TECHNICIAN CONFIRMED THAT THE EO STERILIZER WAS CORRECTLY INSTALLED AND OPERATING TO STERIS SPECIFICATIONS. WHEN THE 3RD PARTY TECHNICIAN REMOVED THE EXHAUST MANIFOLD, HE DID NOT FOLLOW THE LOCK OUT/TAG OUT SAFETY PROCEDURES, RESULTING IN THE USER FACILITY OPERATING THE SECOND EO STERILIZER WHICH SHOULD NOT HAVE BEEN OPERATED WHILE THE FIRST UNIT WAS UNDER MAINTENANCE.