MalfunctionILM
SYSTEM 2000
Received Jul 25, 2013 · Event occurred Jul 15, 2013
Report 1419652-2013-00205 · MDR key 3263648
Device
Manufacturer
Arjo Hospital Equipment AbModel number
AR31011-EUProduct problems
- Temperature Problem
Patient
Not reported
- No Patient Involvement
Narrative
Description of Event or Problem
THE ISSUE OCCURRED IN A S2000 BATH, A BATH SYS FOR ASSISTED BATHING AND IT WAS REPORTED FROM THE ARJOHUNTLEIGH TECH THAT HAS CHECKED THE DEVICE "IT WAS INDICATED BY A HI LEGIONELLA VALVE. ROOT CAUSE: TOO LOW TEMPERATURE ON WATER ON CUSTOMER SITE. THE HYGIENE DEPARTMENT OF THE CUSTOMER IS INVOLVED WORKS ON A SOLUTION BY FLUSHING SCHEDULES ETC. HOT WATER FLUSH EXECUTED." (B)(4).