inforMED
MalfunctionILM

SYSTEM 2000

Received Jul 25, 2013 · Event occurred Jul 15, 2013

Report 1419652-2013-00205 · MDR key 3263648

Device

Model number

AR31011-EU

Product problems

  • Temperature Problem

Patient

Not reported

  • No Patient Involvement

Narrative

Description of Event or Problem

THE ISSUE OCCURRED IN A S2000 BATH, A BATH SYS FOR ASSISTED BATHING AND IT WAS REPORTED FROM THE ARJOHUNTLEIGH TECH THAT HAS CHECKED THE DEVICE "IT WAS INDICATED BY A HI LEGIONELLA VALVE. ROOT CAUSE: TOO LOW TEMPERATURE ON WATER ON CUSTOMER SITE. THE HYGIENE DEPARTMENT OF THE CUSTOMER IS INVOLVED WORKS ON A SOLUTION BY FLUSHING SCHEDULES ETC. HOT WATER FLUSH EXECUTED." (B)(4).