inforMED
MalfunctionGXJ

KERRISON DET130 DEG UP 1MM 180MMTHP

Received Jul 3, 2013 · Event occurred Apr 12, 2013

Report 2916714-2013-00073 · MDR key 3234477

Device

Model number

FK906R

Catalog number

FK906R

Product problems

  • Break
  • Detachment of Device or Device Component

Patient

Not reported

  • No Consequences Or Impact To Patient

Narrative

Description of Event or Problem

KERRISON BROKE OFF AT TIP DURING ACDF. TIP WAS FOUND IN PT. NO PT INJURY OR PROLONGING OF SURGERY.

Additional Manufacturer Narrative

EVAL: THE INSTRUMENT WAS MICROSCOPICALLY CHECKED. THE BREAKAGE AREA OF THE BROKEN FOOT PLATE FOLLOWED A PATTERN KNOWN AS "TRANSCRYSTALLINE FORCE FRACTURE". THE DIRECTION OF BREAKAGE IS TO THE RIGHT FRONT. ANY HINTS FOR A MATERIAL OR MFG ISSUE COULD NOT BE FOUND. BASED ON THE INVESTIGATION RESULTS, THIS ISSUE IS LIKELY RELATED TO HANDLING INFLUENCES.