MalfunctionILM

SYSTEM 2000

Received Jul 11, 2013 · Event occurred Jun 18, 2013

Report 1419652-2013-00192 · MDR key 3234045

Device

Model number

AR

Product problems

  • Defective Device

Patient

Not reported

  • No Known Impact Or Consequence To Patient

Narrative

Description of Event or Problem

THE ISSUE OCCURRED IN A SYSTEM 2000, A BATH SYS FOR ASSISTED BATHING AND IT WAS REPORTED THAT: "FOUND TUB IN HIGHEST POSITION. TURNED POWER ON TO TUB, CAN HEARD HYDRAULIC PUMP RUNNING. SHUT OFF POWER TUB. CHECKED PUSH BUTTONS AND ELECTRICAL SWITCHES ON PANEL, NO ISSUES. CHECKED POWER BOARD, TURNED POWER BACK ON, TAPPED RELAYS ON BOARD, HYDRAULIC PUMP STOPPED RUNNING. LOWERED TUB, TRIED UP BUTTON AGAIN AND THE RELAY DID NOT RELEASE WHEN I RELEASED THE BUTTON. REPLACED PCB, TESTED, NO ISSUES. NO INJURIES WERE REPORTED DUE TO THIS ISSUE." REF #9611530-2013-00083.