MalfunctionILM
SYSTEM 2000
Received Jul 11, 2013 · Event occurred Jun 18, 2013
Report 1419652-2013-00192 · MDR key 3234045
Device
Manufacturer
Arjo Hospital Equipment AbModel number
ARProduct problems
- Defective Device
Patient
Not reported
- No Known Impact Or Consequence To Patient
Narrative
Description of Event or Problem
THE ISSUE OCCURRED IN A SYSTEM 2000, A BATH SYS FOR ASSISTED BATHING AND IT WAS REPORTED THAT: "FOUND TUB IN HIGHEST POSITION. TURNED POWER ON TO TUB, CAN HEARD HYDRAULIC PUMP RUNNING. SHUT OFF POWER TUB. CHECKED PUSH BUTTONS AND ELECTRICAL SWITCHES ON PANEL, NO ISSUES. CHECKED POWER BOARD, TURNED POWER BACK ON, TAPPED RELAYS ON BOARD, HYDRAULIC PUMP STOPPED RUNNING. LOWERED TUB, TRIED UP BUTTON AGAIN AND THE RELAY DID NOT RELEASE WHEN I RELEASED THE BUTTON. REPLACED PCB, TESTED, NO ISSUES. NO INJURIES WERE REPORTED DUE TO THIS ISSUE." REF #9611530-2013-00083.