Received Jul 1, 2013 · Event occurred Apr 29, 2013

Report 1217157-2013-00115 · MDR key 3215905

Device

Generic name

Rapidpoint 500

Catalog number

10492730

Product problems

  • Component Missing

Patient

Not reported

  • No Known Impact Or Consequence To Patient

Narrative

Description of Event or Problem

CUSTOMER REPORTED THAT THE INSTRUMENT MISSED SOME DIGITS ON THE PT ID BARCODE SCAN WHEN THE RESULTS WERE SENT TO THE DATA MANAGEMENT SOFTWARE (RAPIDCOMM). THERE WAS NO REPORT OF INJURY FOR THIS EVENT.

Additional Manufacturer Narrative

THE CUSTOMER HAS BEEN INSTRUCTED TO PROGRAM A SAMPLE LABEL IN THE BARCODE SET UP USING THE PT ID CONFIGURATION. THE INSTRUMENT IS OPERATING AS INTENDED.