inforMED
InjuryEHO

DAUTREY ARCH BAR, 15CM

Received Jul 3, 2013 · Event occurred Jun 10, 2013

Report 0008010177-2013-00149 · MDR key 3205585

Device

Generic name

Arch Bar

Catalog number

01-02960

Lot number

6000008880

Product problems

  • Scratched Material

Patient

Not reported

  • No Known Impact Or Consequence To Patient

Narrative

Description of Event or Problem

IT WAS REPORTED THAT THE ARCH BAR HAS EDGES. IT PULLS OR TEARS GUMS. THE RAIL IS VERY EDGY. THE CUSTOMER USED A DIFFERENT PRODUCT.

Additional Manufacturer Narrative

INVESTIGATION IN PROGRESS BUT NOT YET COMPLETE.

Additional Manufacturer Narrative

THE REPORTED EVENT COULD BE CONFIRMED. VISUAL INSPECTION REVEALED THAT THE EDGES OF THE CURRENT ARCH BARS ARE VERY SHARP. HOWEVER, WITHIN FUNCTIONAL INSPECTION THE REPORTED EVENT COULD NOT BE REPRODUCED. ALSO A COMPARISON WITH THE CURRENT DRAWING HAS SHOWN THAT THE PRODUCTS MEET THE SPECIFICATION. TO ADDRESS THE CUSTOMER CONCERNS A CHANGE REQUEST WAS INITIATED ON 2014-MAR-10 TO REDESIGN THE ARCH BAR IN A WAY IT GETS SMOOTHER. AS A PART OF THE REDESIGN THE EDGES WILL BE NEW DEFINED (ROUNDED OFF) AND THE PEGS MORE CURVED. THE CHANGE REQUEST IS STILL ONGOING.

Description of Event or Problem

IT WAS REPORTED THAT THE ARCH BAR HAS EDGES. IT PULLS OR TEARS GUMS. THE RAIL IS VERY EDGY. THE CUSTOMER USED A DIFFERENT PRODUCT.