inforMED
MalfunctionIXA

AFP

Received Mar 12, 2001

Report MW4002959 · MDR key 320254

Device

Generic name

Soft Film Cassette, 6x12" For Panoramic X-ray System

Manufacturer

Afp Imaging Corp

Model number

*

Catalog number

857000001

Lot number

*

Product problems

  • Design/structure problem
  • Leak/Splash

Patient

NA

Narrative

Description of Event or Problem

SOFT FILM CASSETTES FOR PANORAMIC X-RAY SYSTEMS ARE NOT LIGHT TIGHT. X-RAY FILM IS FOGGED WITHIN 5 MINS WHEN CASETTE IS REMOVED FROM THE DARKROOM, COMPLAINANT HAS WORKED WITH SUPPLIER AND HAS DETERMINED THAT 10 UNITS ALL HAD THE SAME PROBLEM. AT THE REQUEST OF AFP, A COMPARISON WAS MADE WITH A COMPETITOR'S (GENDEX) PRODUCT; THE GENDEX PRODUCT WAS LIGHT TIGHT. AFP CASSETTE IS USED WITH LANEX REGULAR SCREENS AND E-MAT-G FILM. THE 6/12" SIZE IS USED ON A KAYKOR PANARAMIC X-RAY SYSTEM.