inforMED
InjuryHJM

LIGHTMAT

Received Jun 26, 2013 · Event occurred Jun 14, 2013

Report MW5030716 · MDR key 3202300

Device

Generic name

Surgical Illuminator

Manufacturer

Lumitex

Model number

UA2550

Catalog number

UA2550

Lot number

I02386

Product problems

  • Overheating of Device

Patient

48 YR

  • Burn(s)

Narrative

Description of Event or Problem

PT IN THE OPERATING ROOM FOR BILATERAL MASTECTOMY AND RECONSTRUCTION. THE LIGHTMAT, SURGICAL ILLUMINATOR, WAS USED THROUGH OUT THE PROCEDURE. DURING THE PROCEDURE, THE LIGHT CORD SAME IN CONTACT WITH PT CAUSING 3 DIFFERENT BURNS. THE LIGHTMAT AS USES WITH THE RECOMMENDED LIGHT CORD FROM LUMITEX AND A STRYKER X8000 LIGHT SOURCE AT 100%. IT WAS NOTED THAT THE METAL CONNECTION PIECE BETWEEN THE LIGHTMAT AND THE LIGHT CORD BECAME VERY HOT.