FRESENIUS DIALYSIS DELIVERY SYSTEM

Received Mar 9, 2001 · Event occurred Feb 9, 2001

Report 2937457-2001-00004 · MDR key 319854

Device

Generic name

Hemodialysis Machine

Model number

2008K

Catalog number

NA

Lot number

NA

Product problems

  • Use of Incorrect Control/Treatment Settings
  • Display or Visual Feedback Problem

Patient

81 YR

  • Low Blood Pressure/ Hypotension
  • Nausea
  • Vomiting
  • Other (for use when an appropriate patient code cannot be identified)
  • Hospitalization required

Narrative

Description of Event or Problem

A DIALYSIS FACILITY REPORTED THAT THREE PATIENTS COMPLAINED OF NAUSEA, VOMITING AND BECAME HYPOTENSIVE AFTER DIALYZING ON THE SAME MACHINE. THE PATIENTS WERE ADMITTED TO THE HOSPITAL WITH A DIAGNOSIS OF METABOLIC ACIDOSIS. INITIAL INFORMATION RECEIVED WAS THAT PH AND CONDUCTIVITY WERE REPORTEDLY NORMAL. ADDITIONAL INFORMATION RECEIVED LATER VERIFIED THAT RO WATER PH WAS NORMAL BUT DIALYSATE PH WAS NOT CHECKED. NO MACHINE ALARMS WERE REPORTED FRESENIUS EQUIPMENT SPECIALIST FOUND THE MACHINE WITH INCORRECT SODIUM AND BICARBONATE SETTING. BICARBONATE VALUE WAS UNKNOWN SINCE THE MACHINE SHOWED AN ERRONEOUS DISPLAY. THE PATIENTS WERE IN THE HOSPITAL FOR ABOUT A WEEK AND HAVE BEEN DISCHARGED SINCE.