inforMED
MalfunctionGDZ

HOLDER F/VISIOS+FRA TRIAL-IMPL

Received Jun 17, 2013 · Event occurred Nov 10, 2011

Report 8030965-2013-03230 · MDR key 3173890

Device

Manufacturer

Synthes Gmbh

Catalog number

397.089

Lot number

3233260

Product problems

  • Break
  • Sticking

Patient

Not reported

  • No Information

Narrative

Additional Manufacturer Narrative

SYNTHES IS SUBMITTING THIS REPORT AS A RESULT OF REMEDIATION ACTIVITIES RELATED TO FDA WARNING LETTER DATED FEBRUARY 2012. DEVICE(S) LISTED IN THIS REPORT IS (ARE) USED FOR TREATMENT, NOT DIAGNOSIS. ANY ADDITIONAL INFORMATION RECEIVED REGARDING THIS EVENT AFTER FILING THIS REPORT SHALL BE FILED ON A SUPPLEMENTAL MDR. THE ADDITIONAL EVALUATION REVEALED THAT THE THREAD PART OF THE INNER SHAFT IS BROKEN. THE THREAD PART IS STUCK INTO VISIOS TRIAL IMPLANT. NO MANUFACTURING RELATED CONDITIONS WERE FOUND AND WE HAVE TO ASSUME THAT THE INSTRUMENT WAS MOVED IN CRANIAL AND CAUDAL DIRECTIONS. THE DESIGN OF THE HOLDER FOR VISIOS WAS PLANNED FOR LATERAL MOVEMENTS TO REDUCE THE FORCES WHICH WERE CAUSED ON THE INNER SHAFT. HOWEVER, A REVIEW OF THE DEVICE HISTORY RECORDS HAS BEEN REQUESTED.

Description of Event or Problem

IT WAS REPORTED THAT THE INNER PART OF 397.089 BROKE WHILE IT WAS BEING PLACED IN THE TRIAL IMPLANT 396.453. THE TIP OF THE INNER PART OF 397.089 IS NOW STUCK IN THE TRIAL IMPLANT. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).

Additional Manufacturer Narrative

SYNTHES IS SUBMITTING THIS REPORT AS A RESULT OF REMEDIATION ACTIVITIES RELATED TO FDA WARNING LETTER DATED FEBRUARY 2012. DEVICE(S) LISTED IN THIS REPORT IS (ARE) USED FOR TREATMENT, NOT DIAGNOSIS. ANY ADDITIONAL INFORMATION RECEIVED REGARDING THIS EVENT AFTER FILING THIS REPORT SHALL BE FILED ON A SUPPLEMENTAL MDR. A REVIEW OF THE DEVICE HISTORY RECORDS WAS PERFORMED AND NO COMPLAINT RELATED ISSUES WERE FOUND. PLACEHOLDER.