AMSORB 1K CARTRIDGES
Received Dec 19, 2000 · Event occurred Nov 1, 2000
Report 1411365-2000-00004 · MDR key 309579
Device
Generic name
Carbon Dioxide Absorbent
Manufacturer
Armstrong Medical Ltd.Model number
NACatalog number
12344
Lot number
67-009-DD
Product problems
- Absorption
- Other (for use when an appropriate device code cannot be identified)
Patient
UNKNOWN
- Manual ventilation required
Narrative
Description of Event or Problem
REPORT RECEIVED OF INADEQUATE CARBON DIOXIDE ABSORPTION DURING ANESTHESIA. THE CUSTOMER STATES, "THAT AFTER THE AMSORB HAS BEEN USED FOR SEVERAL SURGERIES, THEY START TO GET INCREASED CO2 LEVELS. THEY SHAKE THE CANISTERS AND GET ABOUT ANOTHER FIVE MINUTES OUT OF THE AMSORB, BEFORE THE CO2 LEVEL BEGINS TO RISE AGAIN." WHEN THIS OCCURS THEY CHANGE OUT THE AMSORB FOR BARALYME. THE CUSTOMER REPORTS THAT THE AMSORB IS "CHANNELING AND CHANGES COLOR IN THE CENTER FROM TOP TO BOTTOM." THE COLOR INDICATOR IS ADDED DURING MANUFACTURE AND HAS A SENSITIVE PH FACTOR THAT CAUSES THE GRANULES TO CHANGE COLOR AS THEY NEAR EXHAUSTION. ALL OF THE PTS ARE MECHANICALLY VENTILATED WITH FLOW RATES AT 1-2L/MIN. THE CUSTOMERS USES "STANDARD" ANESHTESIA MACHINES. NO ADVERSE PT EFFECT HAS BEEN REPORTED. ADD'L INFO WAS REQUESTED, BUT NO FURTHER INFO WAS AVAILABLE.