inforMED
InjuryIZW

EUREKA LINEAR III

Received Nov 8, 2000 · Event occurred Oct 17, 2000

Report 1423380-2000-00001 · MDR key 304031

Device

Generic name

Collimator

Manufacturer

Progeny, Inc.

Model number

LINEAR III

Catalog number

70-50000

Lot number

*

Product problems

  • Hole In Material
  • Material Separation

Patient

UNKNOWN

  • Laceration(s)
  • Injury
  • Therapy/non-surgical treatment, additional

Narrative

Description of Event or Problem

COLLIMATOR FELL FROM TUBESTAND DURING A TOMOGRAPHIC EXAMINATION, STRIKING THE PT IN THE FACE. LACERATIONS REQUIRED STITCHES FOR TREATMENT. HOSP REPORTED THREADED HOLE IN MOUNTING COLLAR TO BE STRIPPED. PRODUCT HAS NOT BEEN MADE AVAILABLE TO MFR FOR ANALYSIS.

Remedial action

  • Modification/Adjustment