inforMED
OtherIXY

DUAL HEAD GENESYS

Received Oct 30, 2000 · Event occurred Oct 16, 2000

Report 2916556-2000-00358 · MDR key 303260

Device

Generic name

Nuclear Medicine Equipment

Manufacturer

Adac Laboratories

Model number

2144-3000A

Catalog number

NA

Lot number

NA

Product problems

  • Migration or Expulsion of Device
  • Program, failure to

Patient

Not reported

  • Device induced
  • Injury

Narrative

Description of Event or Problem

IT WAS REPORTED TO ADAC THAT AFTER THE COLLISION SENSOR STOPPED THE GANTRY MOVEMENT, SYSTEM COULD NOT BE PLACED INTO OVERRIDE AND MOVED AWAY DUE TO THE CONDITION OF THE PATIENT. HOSPITAL PERSONNEL CONTACTED ADAC AND WERE INSTRUCTED TO SHUT THE SYSTEM OFF AND MANUALLY ROLL THE GANTRY AWAY. WHILE PUSHING THE GANTRY AWAY, PATIENT'S CENTER LINE WAS INADVERTENTLY PULLED OUT. HOSPITAL PERSONNEL REPORTED THAT THE PATIENT'S CONDITION STABLIZED WITHIN 10 MINUTES.

Remedial action

  • Other