inforMED
OtherILK

SUMMIT STAIRWAY LIFT

Received Jul 5, 2006 · Event occurred Jun 28, 2006

Report 3004153586-2006-00001 · MDR key 3031165

Device

Generic name

Patient Powered Transport Device-stair Lift

Manufacturer

Summit Lifts, Inc.

Model number

AC UNIT

Product problems

  • Disassembly
  • Use of Device Problem

Patient

70 YR

  • Injury

Narrative

Description of Event or Problem

"HE STATED LIFT WAS INSTALLED (B)(6) 2004, AND HAD BEEN USED INFREQUENTLY UNTIL RECENTLY. HE DISASSEMBLED LIFT TO DETERMINE WHY THE SLACK CABLE SAFETY DEVICE WAS PREVENTING LIFT FROM OPERATING FREELY. HE SAID SAFETY DEVICE WAS SETTING WHETHER OR NOT WEIGHT WAS ON THE UNIT. HE TOOK THE LIFT OUT OF THE TRACK AFTER REMOVING THE TOP 2' SECTION OF TRACK. HE CLEANED THE TRACK, AND MODIFIED THE SAFETY DEVICE BY TURNING THE CONE POINT ALLEN SCREW HALF TURN INWARD. HE THEN ATTEMPTED TO PUT THE LIFT BACK TOGETHER TO SEE IF IT WOULD WORK BETTER. (MY NOTES: LIFTS ARE SENT OUT ON 2' SECTIONS OF TRACK FROM THE FACTORY. I TOLD HIM, THE SAFETY DEVICES ARE CALIBRATED, LOCTITE, AND ARE NOT INTENDED TO BE ADJUSTED IN THE FIELD UNDER ANY CIRCUMSTANCES). IN TESTING THE LIFT, HE SAT ON IT TO RIDE IT DOWN. THE LIFT WENT DOWN THE TRACK AT A RATE SUFFICIENT ENOUGH TO DAMAGE THE END OF THE TRACK, EJECTING HIM FROM THE CHAIR AND RESULTING IN HIM BREAKING A BONE IN HIS SHOULDER. I ASKED HIM HOW HE THOUGHT THAT HAPPENED BECAUSE THE SAFETY DEVICE IS DESIGNED TO PREVENT THIS TYPE OF INCIDENT. HE SAID HE DID SOMETHING "STUPID" WHEN PUTTING THE LIFT BACK TOGETHER. IN SOME MANNER, HE HAD ASSEMBLED THE LIFT BACK TO THE TRACK, BUT DIDN'T GET THE UPHILL SET OF WHEELS IN THE TRACK, BUT DIDN'T GET THE UPHILL SET OF WHEELS IN THE TRACK CHANNEL PROPERLY, ALLOWING THEM TO RIDE ABOVE THE TRACK, THEREBY ELIMINATING THE SAFETY DEVICE FROM ENGAGING.

Additional Manufacturer Narrative

USER STATED HEADER ASSEMBLY WENT DOWN WITH THE LIFT. (MY NOTES: IN HOUSE TESTING INDICATES THE HEADER WOULD REMAIN IN PLACE IF ASSEMBLED TO THE TRACK CORRECTLY - EVEN IF THE SCREWS WERE LEFT OUT. THEREFORE, IT WOULD APPEAR THE HEADER WAS NOT PUT ON THE END OF THE TRACK CORRECTLY OR NOT AT ALL). HE ASKED WHAT WE COULD DO TO GET THE LIFT WORKING PROPERLY, STATING HIS SON HAD COME OVER AND REASSEMBLED THE LIFT TO THE TRACK. I TOLD HIM HE COULD SHIP THE LIFT BACK TO OUR FACILITY AND WE WOULD BE GLAD TO GO THROUGH THE UNIT 100% TO MAKE SURE ALL COMPONENTS WERE IN WORKING ORDER. HE DECLINED THIS OFFER. HE SAID HE USED A PIECE OF METAL TO DISABLE THE SAFETY DEVICE SO HIS WIFE COULD USE THE LIFT. I STRONGLY ADVISED HIM AGAINST THIS TYPE OF MODIFICATION, STATING THAT I COULD NEVER RECOMMEND THE USE OF ANY OF OUR LIFTS WITH ANY SAFETY DEVICES DISABLED. HE SAID HE UNDERSTOOD, BUT THEN WENT ON TO ATTEMPT TO GET ME TO OK USE OF THE LIFT, ASKING ABOUT THE LIKELIHOOD OF A GEAR FAILING, DESIGN OF THE GEAR BOX, ETC. I REITERATED, HE NOR ANYONE ELSE SHOULD NOT USE THE LIFT WITH THE SAFETY DEVICE DISABLED. I PROVIDED HIM THE NAME OF A LOCAL DEALER TO COME OUT TO LOOK AT THE LIFT TO DETERMINE IF THE LIFT COULD BE PUT BACK INTO PROPER WORKING ORDER. HE SAID HE DIDN'T WANT TO PAY TOO MUCH FOR THIS SERVICE. I STATED THE DEALER IN HIS AREA HAD A GOOD WORKING UNDERSTANDING OF LIFTS AND STRONGLY ADVISED HE HAVE HIM COME TO HIS HOME AND CALL ME SO I COULD OBTAIN ADDITIONAL INFORMATION TO MAKE A JUDGEMENT ABOUT THE CONDITION OF THE EQUIPMENT." WE HAVE NO RECORD OF (B)(6) OR ANYONE ELSE CALLING US ABOUT ANY PROBLEMS WITH THIS LIFT UNTIL THIS EVENT WAS REPORTED.

Remedial action

  • Repair