UNICEL® DXH SLIDEMAKER STAINER COULTER® CELLULAR ANALYSIS SYSTEM

Received Mar 18, 2013 · Event occurred Feb 17, 2013

Report 1061932-2013-00342 · MDR key 3008064

Device

Generic name

Spinner, Slide, Automated

Manufacturer

Beckman Coulter

Catalog number

775222

Lot number

NA

Product problems

  • Fluid/Blood Leak

Patient

Not reported

  • No Known Impact Or Consequence To Patient

Narrative

Description of Event or Problem

A CUSTOMER REPORTED TO BECKMAN COULTER (BEC) THAT THE UNICEL DXH SLIDEMAKER STAINER COULTER CELLULAR ANALYSIS SYSTEM LEAKED APPROXIMATELY 3 ML OF CLEAR FLUID DURING SHUTDOWN. THE LEAK WAS NOT CONTAINED WITHIN THE INSTRUMENT. THE OPERATOR WAS WEARING GLOVES AND A LAB COAT AT THE TIME OF THE EVENT. THERE WAS NO INJURY OR EXPOSURE TO OPEN WOUNDS OR MUCOUS MEMBRANES, AND THE OPERATOR DID NOT SEEK MEDICAL ATTENTION. MSDS WAS NOT REVIEWED, BUT THE FACILITY HAS AN EXPOSURE CONTROL PLAN. NO ERRONEOUS RESULTS WERE GENERATED.

Additional Manufacturer Narrative

BEC FIELD SERVICE ENGINEER (FSE) OBSERVED THAT THE SOURCE OF THE LEAK WAS THE WASH COLLAR WHICH NEEDED ALIGNMENT. THE FSE PERFORMED THE WASH COLLAR ALIGNMENT FOLLOWED BY DAILY CHECKS AS WELL AS SHUTDOWNS. NO FURTHER EVIDENCE OF LEAKING WAS OBSERVED. (B)(4).