inforMED
MalfunctionKXN

COROMETRICS

Received Sep 22, 2000

Report MW1020004 · MDR key 297738

Device

Generic name

Fetal Monitor

Model number

COROMETRICS 118

Catalog number

*

Lot number

*

Product problems

  • False Device Output

Patient

*

Narrative

Description of Event or Problem

ADD'L INFO REC'D FROM MFR 11/13/00: IN PERFORMING A RISK ASSESSMENT, IT WAS DETERMINED THAT, IN ORDER, FOR DATA TO BE SENT TO THE WRONG BED, THREE SIMULTANEOUS EVENTS MUST OCCUR -(1) THE MONITORS MUST BE ON A COROLAN NETWORK; (2) ADDRESS CHECKING IN THE MONITORS / WALLPLATES IS NOT TURNED ON; AND (3) A HARDWARE FAILURE IN THE COROLAN OR THE CABLES MUST OCCUR. SINCE THE LIKELIHOOD OF ALL THREE EVENTS OCCURRING SIMULTANEOUSLY IS SMALL, THE PROBABILITY OF OCCURRENCE IS NEGLIGIBLE. AS DISCUSSED ABOVE, THE HOSP QS SYSTEM HAS BEEN INSTALLED FOR OVER 5 YEARS AND PREVIOUSLY HAD NOT REPORTED PROBLEMS OF THIS NATURE. IT IS LIKELY THAT STRESSES WERE PLACED ON THE CABLES OR MONITORS, CAUSING A HARDWARE FAILURE (OPEN ADDRESS LINES). MFR'S INFO TECHNOLOGIES' SVC WENT TO THE FACILITY TO VERIFY THAT THE MONITORS, CABLES, WALLPLATES, ETC. WERE WORKING PROPERLY WHILE THE MONITOR AND COROLAN NETWORK CONTINUED TO BE USED AT HOSP. NO PROBLEMS WERE FOUND. IN RESPONSE TO THE CUSTOMER REQUEST, SVC HAS COMPLETED THE PROCESS OF REPLACING THE COROLAN NETWORK WITH A SERIAL MONITORING NETWORK.

Description of Event or Problem

DATA FROM THE FETAL MONITORS IS BEING INCORRECTLY PLACED IN THE ELECTRONIC CHARTING SYSTEM. DATA IS BEING TRANSPOSED FROM ONE PT TO ANOTHER - E.G., BED #1 DATA MAY BE INCORRECTLY ASSIGNED TO BED #2.