inforMED
MalfunctionFAZ

CRYOTECH LCS

Received Jan 18, 1996 · Event occurred Dec 12, 1995

Report 1218402-1996-00005 · MDR key 29695

Device

Generic name

Cryosurgical Device

Catalog number

LCS 3000

Product problems

  • Component(s), overheating of
  • Electrical /Electronic Property Problem
  • Device Emits Odor

Patient

*

  • No Consequences Or Impact To Patient

Narrative

Description of Event or Problem

WHILE PREPARING FOR A CRYOSURGICAL PROCEDURE THE STAFF DETECTED A BURNING SMELL. THE SYSTEM WAS SHUT DOWN AND MOVED TO A ROOM WITH AN OUTSIDE VENT HOOD. UPON EXAMINATION IT WAS NOTED THE EXHAUST HEAT EXCHANGER HAD MELTED. NO PTS WERE IN THE ROOM AT THE TIME. THERE WAS NO KNOWN STAFF INJURY.