MalfunctionFSO
S7 CEILING MOUNT/RIGID COLUMN FOR OPMI PROERGO
Received Jan 11, 2013 · Event occurred Dec 18, 2012
Report 9615010-2013-00001 · MDR key 2918802
Device
Generic name
Microscope Surgical, General & Plastic...
Manufacturer
Carl Zeiss Meditec Ag (oberkochen Site)Model number
S7 CEILING MOUNTCatalog number
000000-1103-022
Product problems
- Device Slipped
Patient
Not reported
- No Known Impact Or Consequence To Patient
Narrative
Additional Manufacturer Narrative
THE FIELD SERVICE REPRESENTATIVE DISCOVERED THE LOOSE SCREWS DURING A SERVICE VISIT. SHE SUBSEQUENTLY TIGHTENED THE LOOSE SCREWS TO SPECIFICATIONS. SITE CONTACT: (B)(6).
Description of Event or Problem
ALL FOUR OF THE SOCKET HEAD CAP SCREWS THAT SECURE THE S7 SUSPENSION SYSTEM CARRIER ARM WITH OPMI PROERGO DENTAL MICROSCOPE TO THE RIGID CEILING COLUMN MOUNT AT THE USER SITE WERE FOUND TO BE LOOSE.
Remedial action
- Modification/Adjustment