inforMED
OtherHSH

3D KNEE

Received Dec 12, 2012 · Event occurred Nov 27, 2012

Report 1644408-2012-00650 · MDR key 2875442

Device

Generic name

Dcm Insert, 9mm

Catalog number

391-09-612

Lot number

53929018

Product problems

  • Loose or Intermittent Connection

Patient

82 YR

  • No Code Available

Narrative

Description of Event or Problem

REVISION SURGERY - THE SCREW IN THE POLYETHYLENE OF THE PT'S KNEE PROSTHESIS HAD LOOSENED AND NEEDED TO BE SWAPPED FOR A NEW INSERT.

Additional Manufacturer Narrative

THE REASON FOR THIS REVISION SURGERY WAS TO REMEDY A LOOSE SCREW FOR A POLY INSERT AFTER 4.3 YEARS OF PATIENT USE. THE OUTCOMES ATTRIBUTED TO THIS EVENT ARE REQUIRED INTERVENTION TO PREVENT PERMANENT IMPAIRMENT/DAMAGE. THE HEALTHCARE PROFESSIONAL INDICATED A SIGNIFICANT ADVERSE EVENT TO THE PATIENT. THERE WAS NO DELAY IN SURGERY AND ANOTHER SUITABLE DEVICE WAS AVAILABLE FOR USE. THE REVISION SURGERY WAS COMPLETED AS INTENDED. THE DEVICE WAS NOT RETURNED TO DJO SURGICAL FOR EXAMINATION. A REVIEW OF THE DEVICE HISTORY RECORDS SHOWED NO NON-CONFORMING MATERIAL REPORTS ASSOCIATED WITH THIS PRODUCT. A REVIEW OF THE PRODUCT COMPLAINT REPORT HISTORY SHOWS THIS IS THE FOURTH COMPLAINT FOR THIS PART NUMBER: TWO DUE TO INFECTION AND ONE DUE TO TRAUMA. THIS IS THE FIRST COMPLAINT FOR THIS LOT NUMBER. THE ROOT CAUSE FOR THE LOOSE SCREW AND POLY INSERT WAS NOT DETERMINED WITH CONFIDENCE. THERE IS NO INFORMATION REPORTED THAT SHOWED A MATERIAL, DESIGN, OR MANUFACTURING PROBLEM WITH THE PRODUCT. THERE ARE NO INDICATIONS OF A PRODUCT OR PROCESS ISSUE AFFECTING IMPLANT SAFETY OR EFFECTIVENESS.