inforMED
InjuryHTG

UNKNOWN DEPUY STD+ PATELLA POLY

Received Dec 10, 2012 · Event occurred Nov 30, 2012

Report 1818910-2012-83332 · MDR key 2864692

Device

Generic name

Patella Prosthesis

Catalog number

UNK-KNEE

Lot number

UNKNOWN

Product problems

  • Naturally Worn

Patient

85 YR

  • Unspecified Infection

Narrative

Description of Event or Problem

PATIENT WAS REVISED TO ADDRESS INFECTION AND POLY WEAR OF THE MESCAL BEARINGS AND PATELLA.

Additional Manufacturer Narrative

THE DEVICE ASSOCIATED WITH THIS REPORT WAS NOT RETURNED. REVIEW OF THE DEVICE HISTORY RECORDS AND/OR A COMPLAINT DATABASE SEARCH WAS NOT POSSIBLE AS THE PRODUCT LOT CODE REQUIRED WAS NOT PROVIDED. THE INVESTIGATION COULD NOT DRAW ANY CONCLUSIONS REGARDING THE REPORTED EVENT WITH THE INFORMATION AVAILABLE; HOWEVER, WEAR OF A POLYETHYLENE DEVICE AFTER BEING IMPLANTED FOR APPROXIMATELY TWENTY-YEARS SHOULD NOT BE UNEXPECTED. INFECTION AFTER THE REPORTED PERIOD OF TIME IS UNLIKELY TO BE PRODUCT RELATED. BASED ON THE INABILITY TO IDENTIFY A ROOT CAUSE, THE NEED FOR CORRECTIVE ACTION WAS NOT INDICATED. DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD THE PRODUCT AND/OR ADDITIONAL INFORMATION BE RECEIVED, THE INVESTIGATION WILL BE RE-OPENED.

Additional Manufacturer Narrative

THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.