inforMED
MalfunctionMYR

AMPCOR

Received Jun 29, 2000 · Event occurred Jun 20, 2000

Report 286009 · MDR key 286009

Device

Generic name

Rpr Test Kit

Model number

*

Catalog number

*

Lot number

1138

Product problems

  • Low Readings

Patient

*

  • Misdiagnosis
  • Therapy/non-surgical treatment, additional

Narrative

Description of Event or Problem

SINCE USING THE AMPCOR RPR KIT IN APRIL-MAY 2000 A 30% DECREASE (3.5% TO 2.2%) IN RPR POSITIVITY RATE HAS BEEN OBSERVED. THE LAB HAS RECEIVED COMPLAINTS OF TITERS BEING TOO LOW. THE PARALLEL TEST OF AMPCOR ANTIGEN VS. ASI AND BD ANTIGEN FAILED IN ONE OF THREE LABORATORIES. A 20% DROP (9.25% TO 7.4%) IN CONFIRMED POSITIVES (POSITIVE FTA'S/TOTAL SPECIMENS TESTED) HAS ALSO BEEN OBSERVED FOR THAT TIME PERIOD.