IMPAX CV DISCOMSTORE WITH MEDIA PURGE DAEMON
Received Nov 16, 2012 · Event occurred Jan 28, 2012
Report 1225058-2012-00069 · MDR key 2857540
Device
Generic name
Medical Image Storage Device
Manufacturer
Agfa Healthcare Corp.Model number
ALL VERSIONSCatalog number
NA
Lot number
NA
Product problems
- Device Operates Differently Than Expected
Patient
Not reported
- No Known Impact Or Consequence To Patient
Narrative
Description of Event or Problem
AGFA SUBMITTED MDR REPORT # 1225058-2012-00003 TO THE FDA ON (B)(4) 2012. THIS IS THE 8TH OCCURRENCE BEING REPORTED FOR THE SAME ISSUE/SAME DEVICE, BUT WITH A DIFFERENT CUSTOMER. ON (B)(4) 2012, A AGFA SVC ENGINEER, WHILE PERFORMING A DIFFERENT SVC AT THE CUSTOMER SITE, DISCOVERED THE SITE'S DICOMSTORE WAS MISCONFIGURED, WHICH LED TO MIXED IMAGE STORAGE LOCATIONS. TO MITIGATE THE POTENTIAL FOR DATA LOSS, AGFA SVC HAS APPLIED CONFIGURATION CHANGES TO THE CUSTOMER'S DATABASE TO ADDRESS THE MIXED STORAGE LOCATION ISSUE. THIS ISSUE WAS IDENTIFIED BY AGFA SVC AND NO HARM TO ANY PTS OCCURRED DURING THIS EVENT.
Additional Manufacturer Narrative
MEDIA PURGE DAEMON (MPD) IS AN OLDEN AGFA PRODUCT USED TO PURGE THE IMAGES STORED ON THE PRIMARY MEMORY CACHE ONCE THEY HAVE BEEN ARCHIVED TO A LONG TERM ARCHIVE AND THE AMOUNT OF DATA STORED ON-LINE REACHES A PREDEFINED AMOUNT. MPD ALLOWS THE SYSTEM TO CONTINUOUSLY RECEIVE THE RECENTLY ACQUIRED IMAGES FROM THIRD PARTIES MODALITIES. THIS TOOL WAS DISCONTINUED AND REPLACED BY CARDIO PURGE SVC (CPS) IN (B)(4) 2008. A REPORTABLE CORRECTION IS UNDERWAY FOR THIS ISSUE AND HAS BEEN REPORTED TO THE FDA. FDA REFERENCE # IS Z-2252-2012.
Remedial action
- Repair