inforMED
MalfunctionCBF

CHAMBER, HYPERBARIC

Received Nov 28, 2012 · Event occurred Oct 24, 2012

Report 2020676-2012-00051 · MDR key 2854339

Device

Generic name

Hyperbaric Monoplace Chamber

Model number

3200R

Catalog number

3200R

Product problems

  • Improper Device Output

Patient

Not reported

  • No Known Impact Or Consequence To Patient

Narrative

Description of Event or Problem

DURING SCHEDULED TESTING THE TECH NOTED THAT THE CHAMBER IS COMPRESSING OUTSIDE OF MFR SPECIFICATIONS. SET AT A RATE OF 1, THE CHAMBER IS COMPRESSING AT 1 PSI PER 43 SECONDS. THIS RATE IS FASTER THAN THE SPECIFIED RATE.

Additional Manufacturer Narrative

THE CHAMBER HAD PREVIOUSLY RECEIVED ITS ANNUAL SERVICE MAINTENANCE ON (B)(4) 2012. NO ISSUES REGARDING THE FUNCTIONALITY WERE NOTED ON THE REPORT. EVALUATION OF THE CHAMBER VERIFIED THE REPORTED FAILURE. AT A RATE OF 1 THE CHAMBER WAS COMPRESSING 1PSI PER 44 SECONDS. THE RATE VALVE WAS REPLACED AND THE CHAMBER WAS PERFORMANCE VERIFIED. IT WAS FOUND TO FUNCTION PER FACTORY SPECIFICATION. A REQUEST FOR THE ORIGINAL RATE VALVE HAS BEEN MADE. AN UPDATE WILL BE SENT UPON EVAL OF THE RATE VALVE.