inforMED
MalfunctionKPQ

XKNIFE 5.0 INSTALL KIT

Received Nov 16, 2012

Report 1222895-2012-00048 · MDR key 2845265

Device

Generic name

Neuro

Catalog number

XK5

Lot number

5.0.1

Product problems

  • Device Displays Incorrect Message

Patient

Not reported

  • No Consequences Or Impact To Patient

Narrative

Description of Event or Problem

"X1 AND Y1 JAWS SHOWING AS -3.5 IN (B)(4), SHOULD BE +3.5 DUE TO SOFTWARE ERROR." THE ERROR MESSAGE OCCURS IN THE (B)(4) SYSTEM. THE (B)(4) SYSTEM IS THIRD PARTY SOFTWARE INTEGRATED WITH THE LINAC THAT VERIFIES THE PROPER POSITION OF THE COUCH, COLLIMATOR, GANTRY AND ANY BEAM MODIFIERS BEFORE TREATMENT IS DELIVERED. XKNIFE IS A RADIOSURGERY AND RADIOTHERAPY TREATMENT PLANNING SYSTEM INTENDED FOR USE IN STEREOTACTIC AND NON-STEREOTACTIC (FRAMELESS STEREOTACTIC), COLLIMATED BEAM, COMPUTER PLANNED, LINEAR ACCELERATOR (LINAC) BASED TREATMENT. THE SOFTWARE INSTALLATION IS CONDUCTED PRIOR TO SCHEDULING THE USE OF THE SYSTEM TO TREAT PATIENTS. THEREFORE, THERE WAS NO PATIENT CONTACT.

Additional Manufacturer Narrative

TO DATE, THE DEVICE INVOLVED IN THE REPORTED INCIDENT HAS NOT BEEN RECEIVED FOR EVALUATION. AN INVESTIGATION HAS BEEN INITIATED BASED UPON THE REPORTED INFORMATION.