OHMEDA QUICK CONNECT
Received Sep 21, 2012 · Event occurred Jul 19, 2012
Report 2843524 · MDR key 2843524
Device
Generic name
Flowmeter, Calibration, Gas, Adapter
Manufacturer
Cardinal HealthModel number
*Catalog number
539-2105
Lot number
*
Product problems
- Improper or Incorrect Procedure or Method
- Inaccurate Delivery
- Component or Accessory Incompatibility
Patient
65 YR
- No Information
Narrative
Description of Event or Problem
PATIENT WAS CONNECTED TO MEDICAL AIR AND NOT OXYGEN. WE USE PISS (PIN INDEX SAFETY SYSTEM) FLOWMETERS IN THE NEWER AREAS OF THE HOSPITAL AND DISS(DIAMETER INDEX SAFETY SYSTEM) FLOWMETERS IN THE OLDER WINGS. EITHER WAY YOU MAY NOT PLUG AN OXYGEN FLOWMETER INTO AN AIR OUTLET OR VICE VERSA. THE NIPPLE ADAPTER THAT IS USED TO ATTACH A DELIVERY DEVICE (E.G. MASK, NASAL CANNULA) IS THE SAME SIZE. WE HAVE GREEN/YELLOW COLOR CODED NIPPLES WHICH ARE INTERCHANGEABLE. THE PROBLEM ARISES WHEN STAFF PLACES THE INCORRECT NIPPLE ONTO THE FLOWMETER AND ASSUMES THAT THE NIPPLE COLOR DENOTES THE DELIVERED GAS.