inforMED
MalfunctionMBV

OFFSET TIB TRAY 2.5MM ADAPTOR

Received Nov 20, 2012 · Event occurred Aug 30, 2012

Report 0001825034-2012-02431 · MDR key 2840828

Device

Generic name

Prosthesis, Knee

Manufacturer

Biomet Orthopedics

Model number

N/A

Catalog number

141490

Lot number

378330

Product problems

  • Device Operates Differently Than Expected

Patient

Not reported

  • Unspecified Infection

Narrative

Description of Event or Problem

IT WAS REPORTED THAT PATIENT UNDERWENT REVISION KNEE ARTHROPLASTY PROCEDURE ON (B)(6) 2012 TO REMOVE SPACERS THAT WERE PLACED DUE TO INFECTION. DURING THE PROCEDURE, THE OFFSET TIBIAL TRAY ADAPTOR WOULD NOT ENGAGE. ANOTHER TIBIAL TRAY ADAPTOR WAS AVAILABLE TO COMPLETE THE PROCEDURE WITHOUT INJURY TO THE PATIENT. THERE WAS NO SIGNIFICANT DELAY AS A RESULT.

Additional Manufacturer Narrative

EVALUATION OF RETURNED DEVICE CONFIRMED REPORTED CONDITION. DECISION WAS MADE TO RECALL BASED ON AN OPERATOR ISSUE THAT WAS DETERMINED AS PART OF AN INTERNAL INVESTIGATION. (B)(4).

Remedial action

  • Recall