inforMED
InjuryFML

6020F CHAIR

Received Oct 3, 2012 · Event occurred Sep 10, 2012

Report 3001452523-2012-00001 · MDR key 2774294

Device

Generic name

Blood Drawing Chair

Manufacturer

Clinton Industries

Model number

6020F

Catalog number

6020F

Lot number

NA

Product problems

  • Component Falling

Patient

NI

  • Syncope
  • Fall
  • Head Injury

Narrative

Description of Event or Problem

AS DESCRIBED TO CLINTON INDUSTRIES BY EMAIL FROM (B)(6), END-USER FACILITY, (B)(6): AFTER COMPLETING A BLOOD DRAW FOR RESEARCH PURPOSES AND STILL SEATED IN THE CHAIR WITH THE ARM REST ACROSS HIS BODY, THE RESEARCH PARTICIPANT FAINTED AND FELL FORWARD. AS HE WAS FALLING TO THE FLOOR THE CHAIR WAS TIPPING WITH HIM. THE INDIVIDUAL SUSTAINED SIGNIFICANT FACIAL INJURIES RELATED TO THE FALL. AS DESCRIBED TO CLINTON INDUSTRIES BY EMAIL FROM (B)(6), (DISTRIBUTOR OF DEVICE INVOLVED) DURING A TAPED TELEPHONE INTERVIEW OF "(B)(6)" (LAST NAME INAUDIBLE) FROM (B)(6): PATIENT PASSED-OUT AFTER THE PUNCTURE IN THE ARM, ARM OF THE CHAIR WAS DOWN, THEY HUNCHED DOWN AND THE WHOLE CHAIR FLIPPED FORWARD. THEY FELL ONTO THE GROUND CAUSING SIGNIFICANT INJURIES.

Additional Manufacturer Narrative

PICTURES OF THE CHAIR IN QUESTION WHERE INCLUDED FROM THE END USER AND DID NOT SHOW ANY DEFECTS OR FAILURES.

Remedial action

  • Other