ARTIS ZEE FLOOR MN

Received Sep 24, 2012 · Event occurred Sep 17, 2012

Report 2770156 · MDR key 2770156

Device

Generic name

X-ray System, Cath/angio, With Magnetic Navigation

Model number

10094142

Catalog number

*

Lot number

*

Product problems

  • Device Operates Differently Than Expected

Patient

51 YR

  • No Information

Narrative

Description of Event or Problem

X-RAY IMAGING SYSTEM HAD FAILED AND A VIDEO PC BOARD WAS RE-SEATED ON THE IAS PC. SYSTEM WAS STRESSED MULTIPLE TIMES WITH NO ISSUES.FIVE DAYS LATER, AFTER A PATIENT WAS SEDATED, THE SYSTEM WAS TESTED PRIOR TO THE PROCEDURE AND THE FLUOROSCOPY FUNCTION WAS NOT AVAILABLE. SYSTEM WAS REBOOTED FOUR TIMES WITH NO SUCCESS. THE PROCEDURE WAS CANCELLED. THE REPLACEMENT PC ARRIVED LATER THAT SAME DAY AND WAS REPLACED. SYSTEM IS NOW FULLY OPERATIONAL AFTER A WEEK OF OBSERVATION.