DIMENSION VISTA® CLINICAL CHEMISTRY SYSTEM
Received Aug 30, 2012 · Event occurred Aug 8, 2012
Report 9610806-2012-00011 · MDR key 2720757
Device
Generic name
Immunoglobulin A Flex® Reagent Cartridge
Manufacturer
Siemens Healthcare Diagnostics Products GmbhCatalog number
K7052
Lot number
11195MA
Product problems
- Low Test Results
Patient
Not reported
- No Known Impact Or Consequence To Patient
Narrative
Description of Event or Problem
A FALSELY DEPRESSED IMMUNOGLOBULIN A (IGA) RESULT WAS REPORTED ON A PATIENT SAMPLE. THE RESULT WAS REPORTED TO THE PHYSICIAN WHO QUESTIONED THE RESULT. THE PATIENT WAS RE-RUN ON DILUTION AND AN ELEVATED RESULT WAS OBTAINED. PATIENT TREATMENT WAS NOT ALTERED OR PRESCRIBED ON THE BASIS OF THE FALSELY DEPRESSED IGA RESULT. THERE IS NO REPORT OF ADVERSE OUTCOME TO THE PATIENT AS A RESULT OF THE INITIAL FALSELY DEPRESSED IGA RESULT.
Additional Manufacturer Narrative
ANALYSIS OF THE INSTRUMENT AND INSTRUMENT DATA INDICATE THAT THE CAUSE FOR THE FALSELY DEPRESSED IGA RESULTS IS UNKNOWN. THE DIMENSION VISTA(R) IGA FLEX(R) REAGENT CARTRIDGE IFU IN THE LIMITATIONS OF "PROCEDURE SECTION STATES: ""THE IGA METHOD HAS BEEN DESIGNED TO MINIMIZE ANTIGEN EXCESS IN THE INITIAL SAMPLE DILUTION. HOWEVER, IT CANNOT BE COMPLETELY ELIMINATED AND IN RARE CASES VERY HIGH IMMUNOGLOBULIN CONCENTRATIONS MAY PRODUCE FALSELY-LOW RESULTS. MONOCLONAL IMMUNOGLOBULINS MAY SHOW REACTIVITY DIFFERENT FROM THE POLYCLONAL CALIBRATOR, WHICH IN ISOLATED CASES MAY LEAD TO ARTIFICIALLY DECREASED OR NON-LINEAR RESULTS." THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.