DIMENSION VISTA® CLINICAL CHEMISTRY SYSTEM

Received Aug 30, 2012 · Event occurred Aug 8, 2012

Report 9610806-2012-00011 · MDR key 2720757

Device

Generic name

Immunoglobulin A Flex® Reagent Cartridge

Catalog number

K7052

Lot number

11195MA

Product problems

  • Low Test Results

Patient

Not reported

  • No Known Impact Or Consequence To Patient

Narrative

Description of Event or Problem

A FALSELY DEPRESSED IMMUNOGLOBULIN A (IGA) RESULT WAS REPORTED ON A PATIENT SAMPLE. THE RESULT WAS REPORTED TO THE PHYSICIAN WHO QUESTIONED THE RESULT. THE PATIENT WAS RE-RUN ON DILUTION AND AN ELEVATED RESULT WAS OBTAINED. PATIENT TREATMENT WAS NOT ALTERED OR PRESCRIBED ON THE BASIS OF THE FALSELY DEPRESSED IGA RESULT. THERE IS NO REPORT OF ADVERSE OUTCOME TO THE PATIENT AS A RESULT OF THE INITIAL FALSELY DEPRESSED IGA RESULT.

Additional Manufacturer Narrative

ANALYSIS OF THE INSTRUMENT AND INSTRUMENT DATA INDICATE THAT THE CAUSE FOR THE FALSELY DEPRESSED IGA RESULTS IS UNKNOWN. THE DIMENSION VISTA(R) IGA FLEX(R) REAGENT CARTRIDGE IFU IN THE LIMITATIONS OF "PROCEDURE SECTION STATES: ""THE IGA METHOD HAS BEEN DESIGNED TO MINIMIZE ANTIGEN EXCESS IN THE INITIAL SAMPLE DILUTION. HOWEVER, IT CANNOT BE COMPLETELY ELIMINATED AND IN RARE CASES VERY HIGH IMMUNOGLOBULIN CONCENTRATIONS MAY PRODUCE FALSELY-LOW RESULTS. MONOCLONAL IMMUNOGLOBULINS MAY SHOW REACTIVITY DIFFERENT FROM THE POLYCLONAL CALIBRATOR, WHICH IN ISOLATED CASES MAY LEAD TO ARTIFICIALLY DECREASED OR NON-LINEAR RESULTS." THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.