inforMED
OtherHSH

OPTETRAK

Received Aug 23, 2012 · Event occurred Jul 24, 2012

Report 1038671-2012-00117 · MDR key 2718769

Device

Generic name

Tibial Tray, Cemented, Finned

Manufacturer

Exactech, Inc.

Catalog number

200-04-22

Product problems

  • Naturally Worn

Patient

Not reported

  • No Code Available

Narrative

Description of Event or Problem

REVISION OF KNEE COMPONENTS DUE TO SIGNIFICANT WEAR OF THE POSTERO-MEDIAL ASPECT OF THE INSERT. SURGEON REPORTED THAT PT WAS DOING FINE FROM A CLINICAL POINT OF VIEW AND WAS STABLE ON VARUS AND VALGUS STRESS TEST.

Additional Manufacturer Narrative

ENGINEERING EVAL OF THE RETURNED DEVICE NOTED ADHERENT PMMA BONE CEMENT ON BOTH THE MEDIAL AND LATERAL CEMENT POCKETS AS WELL AS IN THE TIBIAL UNDERCUTS. THE POSTERIOR EDGE OF THE MEDIAL ASPECT SHOWED METAL LOSS PRODUCING A SEMI-CIRCULAR GOUGE EXACTLY CONGRUENT TO THE SHAPE OF THE MEDIAL FEMORAL CONDYLE. THIS IS CONSISTENT WITH ARTICULATION BETWEEN THE TIBIAL TRAY AND FEMORAL CONDYLE. THE DEVICE HISTORY RECORD REVIEW PROVIDES ASSURANCE THE PART WAS ACCEPTED WITH CONFORMANCE TO THE PRODUCT REQUIREMENTS.