inforMED
MalfunctionDTY

TERUMO CDI 100 MONITOR

Received Aug 21, 2012 · Event occurred Jul 26, 2012

Report 1828100-2012-01100 · MDR key 2711116

Device

Model number

100

Catalog number

100

Product problems

  • Display or Visual Feedback Problem

Patient

Not reported

  • No Information

Narrative

Description of Event or Problem

IT WAS REPORTED THAT THE NUMBERS DISPLAYED ON THE MONITOR ARE NO LONGER ACCURATE. NO OTHER DETAILS REGARDING THE NATURE OF THIS EVENT WERE PROVIDED.

Additional Manufacturer Narrative

LAB ANALYSIS FOUND THE UNIT FAILED THE COLOR CHIP TEST AND HAS A DEFECTIVE OPTO PROBE RELATED TO THE AGE OF THE DEVICE. NO PARTS ARE AVAILABLE FOR REPAIR. IF ADDITIONAL INFO BECOMES AVAILABLE ON THIS COMPLAINT THAT WOULD ALTER THE FACTS AND/OR CONCLUSION, A SUPPLEMENTAL REPORT WILL BE FILED ACCORDINGLY.