inforMED
MalfunctionHSH

ITOTAL TIBIAL IMPACTOR TIP

Received Aug 1, 2012 · Event occurred Jul 1, 2012

Report 3004153240-2012-00061 · MDR key 2706501

Device

Generic name

Reusable Tibial Impactor Tip

Manufacturer

Conformis, Inc.

Catalog number

ED-02611

Lot number

M1107

Product problems

  • Break

Patient

60 YR

  • No Information

Narrative

Description of Event or Problem

TIBIAL TRAY IMPACTOR TIP BROKE AT THE POINT WHERE THE TIP CONNECTS TO THE IMPACTOR HANDLE.

Additional Manufacturer Narrative

TIBIAL TRAY IMPACTOR TIP BROKE AT THE POINT WHERE THE TIP CONNECTS TO THE IMPACTOR HANDLE. THE SURGERY WAS COMPLETED SUCCESSFULLY. REVIEW OF THE DEVICE HISTORY RECORD INDICATES THAT THE DEVICE WAS MANUFACTURED TO SPECIFICATION. DEVICE EVAL: THE TIBIAL TRAY IMPACTOR TIP WAS RETURNED TO CONFORMIS FOR EVAL. THE REPORTED FAILURE MODE WAS CONFIRMED. THE POST THAT CONNECTS THE TIP TO THE IMPACTOR HANDLE WAS FOUND TO BE BROKEN OFF INSIDE THE IMPACTOR HANDLE. THE COMPLAINT APPEARS TO BE AN ISOLATED INCIDENT.