inforMED
InjuryMOS

SENSE 7CH 1.5T BREAST COIL

Received Jul 9, 2012 · Event occurred Aug 27, 2010

Report 1056069-2012-00003 · MDR key 2651114

Device

Manufacturer

Invivo

Model number

4522-150-40213

Product problems

  • Insufficient Device Problem Information

Patient

66 YR

  • Bone Fracture(s)

Narrative

Description of Event or Problem

THE PT WAS PREPARED FOR A MAGNETIC RESONANCE (MR) IMAGING BREAST EXAMINATION USING AN IN VIVO 7-CHANNEL BREAST COIL ACCESSORY DEVICE. WHEN THE OPERATOR MOVED THE PT INTO THE MR SCANNER (PHILIPS ACHIEVA 1.5T MODEL), THE PT SCREAMED. THE HEALTH CARE FACILITY REPORTED THAT THE PT SUSTAINED A BROKEN RIB.

Additional Manufacturer Narrative

(B)(4). METHOD/RESULTS: OPERATIONAL CONTEXT CAUSED EVENT. CONCLUSIONS: USER ERROR CAUSED EVENT. THE INJURY WAS ATTRIBUTED TO OPERATOR ERROR DURING MOVEMENT OF THE PT INTO THE BORE OF THE MAGNET SYSTEM. THE INSTRUCTIONS FOR USE PROVIDED FOR THE IN VIVO BREAST COIL DEVICE CAUTIONS THE USER TO ¿ASSURE THAT THE PT IS NOT TOUCHING THE BORE¿.