MalfunctionGKJ

COULTER LH 750 SLIDEMAKER

Received Jun 23, 2012 · Event occurred May 29, 2012

Report 1061932-2012-01959 · MDR key 2626896

Device

Generic name

Spinner, Slide, Automated

Model number

NA

Catalog number

6605633

Lot number

NA

Product problems

  • Hole In Material
  • Moisture Damage

Patient

Not reported

  • No Known Impact Or Consequence To Patient

Narrative

Description of Event or Problem

CUSTOMER REPORTED TO BECKMAN COULTER, INC (BEC) THAT THERE WAS SOME MOISTURE IN THE TRAY UNDERNEATH THE SHUTTLE OF THE COULTER LH 750 SLIDEMAKER. THERE WAS NO REPORT OF PATIENT RESULTS AFFECTED. THERE WAS NO REPORT OF ANY ADVERSE EVENT OR INJURY REQUIRING MEDICAL INTERVENTION OR PATIENT TREATMENT. BEC FIELD SERVICE ENGINEER (FSE) OBSERVED 5-10 ML OF ISOTON IN THE UNIT. THE FSE FOUND A HOLE IN THE BLACK STRIPE I-BEAM TUBING THROUGH PINCH VALVE VL18. THE FSE REPLACED THE DAMAGED TUBING.

Additional Manufacturer Narrative

(B)(4).