RINGLOC SCREWIN PROV +3/HW 36 SZ23

Received Jun 19, 2012 · Event occurred May 22, 2012

Report 0001825034-2012-00883 · MDR key 2621239

Device

Generic name

Template

Manufacturer

Biomet Orthopedics

Model number

N/A

Catalog number

31-108323

Lot number

053360

Product problems

  • Difficult to Remove
  • Device Slipped

Patient

Not reported

  • No Information

Narrative

Description of Event or Problem

IT WAS REPORTED THAT PATIENT UNDERWENT TOTAL HIP ARTHROPLASTY ON (B)(6) 2012. DURING THE PROCEDURE, AS THE ACETABULAR CUP WAS IMPACTED, THE LOCKING RING CAME OUT. AFTER TRIALING, THE TRIAL LINER WAS DIFFICULT TO REMOVE. SUBSEQUENTLY, THE SURGEON INSTALLED A NEW LOCKING RING IN THE ACETABULAR CUP, BUT HAD DIFFICULTIES LOCKING THE ACETABULAR LINER IN PLACE. THE SURGEON USED A DIFFERENT ACETABULAR CUP, LINER, MODULAR HEAD, AND TAPERLOC STEM TO COMPLETE THE PROCEDURE. A DELAY OF MORE THAN HALF AN HOUR OCCURRED.

Additional Manufacturer Narrative

CURRENT INFORMATION IS INSUFFICIENT TO PERMIT A CONCLUSION AS TO THE CAUSE OF THE EVENT. REVIEW OF DEVICE HISTORY RECORDS SHOW THAT LOT RELEASED WITH NO RECORDED ANOMALY OR DEVIATION. EVALUATION IN PROCESS BUT NOT YET COMPLETE. UPON COMPLETION OF EVALUATION, A FOLLOW UP REPORT WILL BE SENT TO THE FDA.

Additional Manufacturer Narrative

VISUAL INSPECTION OF TRIAL SHOWS SIGNS OF DAMAGE ON THE SCALLOPS FROM THE ATTEMPTS AT REMOVAL. THERE ARE ADDITIONALLY TWO AREAS WITH MINOR DAMAGE/SCUFFING IN THE RING GROOVE THAT APPEAR NEAR THE SCALLOP DAMAGE, ALSO PRESUMABLY FROM REMOVAL. DIMENSIONS FOR THE LOCKING RING GROOVE MET SPECIFICATIONS. THE REPORTED EVENT COULD NOT BE SUBSTANTIATED.