MalfunctionLNH
INGENIA 1.5T
Received Jul 22, 2026 · Event occurred Apr 24, 2026
Report 3042177665-2026-000120 · MDR key 25902704
Device
Generic name
Full-body Mri System, Superconducting Magnet
Manufacturer
Philips Medical Systems Nederland B.v.Model number
INGENIA 1.5TCatalog number
781396
Product problems
- Insufficient Device Problem Information
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
PHILIPS HAS STARTED AN INVESTIGATION INTO THIS EVENT. A FOLLOW UP REPORT WILL BE SUBMITTED ONCE COMPLETED.
Description of Event or Problem
PHILIPS RECEIVED A REPORT THAT THE SCAN SEQUENCES WERE ABORTED. AFTER INVESTIGATION IT WAS FOUND THAT THE GRADIENT COIL TERMINAL WAS BURNED AND PARTIALLY BROKEN. BASED ON THIS INFORMATION IT WAS DECIDED TO REPORT THIS EVENT.