inforMED
MalfunctionLNH

INGENIA 1.5T

Received Jul 22, 2026 · Event occurred Apr 24, 2026

Report 3042177665-2026-000120 · MDR key 25902704

Device

Generic name

Full-body Mri System, Superconducting Magnet

Model number

INGENIA 1.5T

Catalog number

781396

Product problems

  • Insufficient Device Problem Information

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

PHILIPS HAS STARTED AN INVESTIGATION INTO THIS EVENT. A FOLLOW UP REPORT WILL BE SUBMITTED ONCE COMPLETED.

Description of Event or Problem

PHILIPS RECEIVED A REPORT THAT THE SCAN SEQUENCES WERE ABORTED. AFTER INVESTIGATION IT WAS FOUND THAT THE GRADIENT COIL TERMINAL WAS BURNED AND PARTIALLY BROKEN. BASED ON THIS INFORMATION IT WAS DECIDED TO REPORT THIS EVENT.