UNITY CATARACT FMS PROCEDURE PACK
Received Jul 22, 2026 · Event occurred Jul 1, 2026
Report 1644019-2026-03942 · MDR key 25901379
Device
Generic name
General Surgery Tray (kit)
Manufacturer
Alcon Research, Llc - HoustonModel number
NACatalog number
8065000339
Lot number
ASKU
Product problems
- Material Separation
Patient
NA · Unknown
- Insufficient Information
Narrative
Additional Manufacturer Narrative
INVESTIGATION INCLUDING ROOT CAUSE ANALYSIS IS IN PROGRESS. A SUPPLEMENTAL MDR WILL BE FILED AS NECESSARY IN ACCORDANCE WITH 21 CFR 803.56 WHEN ADDITIONAL REPORTABLE INFORMATION BECOMES AVAILABLE. THE MANUFACTURER INTERNAL REFERENCE NUMBER IS: (B)(4). H.10 REFLECTS ALL RELATED REPORT NUMBERS ASSOCIATED WITH THIS PRODUCT EVENT THAT HAVE BEEN SUBMITTED AT THIS TIME.
Description of Event or Problem
A HEALTHCARE PROFESSIONAL REPORTED THAT, DURING SURGERY, THE ASPIRATION TUBING OF AN OPHTHALMIC HANDPIECE WAS NOT SECURED WITH A SCREW AND COULD BECOME DETACHED DURING THE CORTEX REMOVAL STAGE. THE PROCEDURE TYPE AND OTHER SURGERY DETAILS WERE UNKNOWN. THERE WAS NO INFORMATION ABOUT PATIENT IMPACT. THIS REPORT PERTAINS TO THE SUCTION TUBING INVOLVED IN THE COMPLAINT. ADDITIONAL INFORMATION HAS BEEN REQUESTED BUT NON-RECEIVED TILL DATE.