inforMED
InjuryKWP

TULIP 5.5/6.0MM HEAD - 7.1MM BALL - STANDARD

Received Jul 21, 2026 · Event occurred Jun 20, 2026

Report 3012447612-2026-00199 · MDR key 25898294

Device

Generic name

Appliance, Fixation, Spinal Interlaminal

Model number

N/A

Catalog number

LS-TU7S5560

Lot number

NI

Product problems

  • Detachment of Device or Device Component

Patient

69 YR · Male

  • Insufficient Information

Narrative

Additional Manufacturer Narrative

D4: THE REMAINDER OF THE UDI NUMBER IS UNKNOWN BECAUSE THE LOT NUMBER IS NOT AVAILABLE. D10: ROD X2, CLOSURE TOPS X6. WITHOUT A PRODUCT RETURN, NO PRODUCT EVALUATION IS ABLE TO BE CONDUCTED. THE LOT NUMBER IS UNKNOWN; THEREFORE, THE DEVICE HISTORY RECORDS ARE UNABLE TO BE REVIEWED. CURRENT INFORMATION IS INSUFFICIENT TO PERMIT A VALID CONCLUSION ABOUT THE CAUSE OF THIS EVENT. IF ADDITIONAL INFORMATION IS OBTAINED THAT ADDS VALUE TO THE RELEVANT CONTENT OF THIS REPORT AND/OR A CONCLUSION CAN BE DRAWN, A FOLLOW-UP REPORT WILL BE SENT. REFERENCE REPORTS 3012447612-2026-00190 .

Description of Event or Problem

IT WAS REPORTED THAT TWO DAYS AFTER A LINESIDER CONSTRUCT WAS INSTRUMENTED AT L3-5, THE PATIENT FELT A POP IN HIS BACK AND A FOLLOW-UP MRI FOUND THE TULIP HAD DETACHED FROM THE SCREW SHANK AT LEFT L3. A REVISION SURGERY WAS PERFORMED TO REPLACE THE SCREW. DURING THE REVISION, THE TULIPS DETACHED FROM THE SHAFTS OF FOUR MORE SCREWS. THE FULL CONSTRUCT WAS REMOVED AND REPLACED WITH COMPETITIVE PRODUCT. THIS IS REPORT TEN OF TEN FOR THIS EVENT.