inforMED
InjuryKWP

DUAL TO QUAD LEAD SHANK 7.1MM BALL 7.5MM X 45MM

Received Jul 21, 2026 · Event occurred Jun 20, 2026

Report 3012447612-2026-00191 · MDR key 25896351

Device

Generic name

Appliance, Fixation, Spinal Interlaminal

Model number

NA

Catalog number

LS-SC7E7545

Lot number

10190004

Product problems

  • Detachment of Device or Device Component

Patient

69 YR · Male

  • Insufficient Information

Narrative

Additional Manufacturer Narrative

D10: ROD X2, CLOSURE TOPS X6. CURRENT INFORMATION IS INSUFFICIENT TO PERMIT A VALID CONCLUSION ABOUT THE CAUSE OF THIS EVENT. A FOLLOW UP REPORT WILL BE SENT UPON COMPLETION OF THE DEVICE EVALUATION. REFERENCE REPORTS: 3012447612-2026-00190 THROUGH 3012447612-2026-00199.

Description of Event or Problem

IT WAS REPORTED THAT TWO DAYS AFTER A LINESIDER CONSTRUCT WAS INSTRUMENTED AT L3-5, THE PATIENT FELT A POP IN HIS BACK AND A FOLLOW-UP MRI FOUND THE TULIP HAD DETACHED FROM THE SCREW SHANK AT LEFT L3. A REVISION SURGERY WAS PERFORMED TO REPLACE THE SCREW. DURING THE REVISION, THE TULIPS DETACHED FROM THE SHAFTS OF FOUR MORE SCREWS. THE FULL CONSTRUCT WAS REMOVED AND REPLACED WITH COMPETITIVE PRODUCT. THIS IS REPORT TWO OF TEN FOR THIS EVENT.