inforMED
InjuryOLO

EXCELSIUS GPS

Received Jul 21, 2026 · Event occurred Jun 16, 2026

Report 3004142400-2026-00146 · MDR key 25895128

Device

Generic name

Robot, Surgical, Navigation Unit

Model number

6143.1001

Lot number

GPS-0253

Product problems

  • Human-Device Interface Problem

Patient

Not reported

  • Spinal Column Injury

Narrative

Additional Manufacturer Narrative

THE DEVICE WAS UNAVAILABLE FOR EVALUATION. NO DETERMINATIONS COULD BE MADE AS TO THE CAUSE OF THE REPORTED ISSUE.

Description of Event or Problem

IT WAS REPORTED THAT DURING AN EXCELSIUSGPS SURGERY, WE SPUN WITH E3D, SCAN LOOKED GREAT, WE PLACED ALL SCREWS, LOW OFFSET, GREEN DRB AND SURVEILLANCE, THEN WE TOOK X-RAYS POST-ROBOT AND LEFT T10/11 WERE MEDIAL INTO THE CANAL. WE REMOVED SCREWS AND THE SURGEON SAW CSF. WE RESPUN, AND REPLACED BOTH SCREWS. THE T11 WAS CORRECTED BUT T10 WENT EXACTLY WHERE IT WAS PREVIOUSLY, AGAIN ALL INDICATORS ON THE ROBOT WERE GREEN. WE REMOVED THE SCREW. THERE WAS ALSO A DURAL TEAR. THE PATIENT WAS DOING WELL THE NEXT DAY WITH NO DEFICITS.