inforMED
InjuryNDP

UNKNOWN ATIS EV TOOL

Received Jul 21, 2026 · Event occurred Jun 9, 2026

Report 3013111692-2026-22716 · MDR key 25893027

Device

Generic name

Accessories, Implant, Dental, Endosseous

Catalog number

UNK ATIS EV TOOL

Product problems

  • Mechanical Problem

Patient

NA · Unknown

  • Failure of Implant

Narrative

Additional Manufacturer Narrative

THEREFORE, BECAUSE A SERIOUS INJURY RESULTED, THIS EVENT IS REPORTABLE PER 21 CFR PART 803. DENTSPLY SIRONA BECAME AWARE OF A SERIOUS INJURY WITH AN IMPLANT DUE TO A MALFUNCTION OF ABUTMENT THAT OCCURRED WHILE USING THE ASTRA TECH IMPLANT SYSTEM EV OS. THIS REPORT IS FOR THE IMPLANT DRIVER DEVICE. THE DENTAL IMPLANT DEVICE REPORT WILL BE SUBMITTED SEPARATELY. PRODUCT RETURN IS REQUESTED AND PRODUCT WILL BE EVALUATED AFTER RECEIPT. IN CASE ANY NEW OR ADDITIONAL INFORMATION WILL BE GAINED FROM THIS INVESTIGATION A FOLLOW-UP REPORT WILL BE SENT. TREND IS TRACKED AND MONITORED.

Description of Event or Problem

IT WAS REPORTED THAT THERE WAS A TOOL MALFUNCTION WHERE THE DRIVER BECAME STRIPPED AND WOULD NOT ALLOW FULL SEATING OF THE IMPLANT NOR REMOVAL. THIS RESULTED IN THE IMPLANT BEING EXPLANTED.