OLYMPUS EVIS EXERA COLONOVIDEOSCOPE
Received May 18, 2012 · Event occurred Apr 1, 2012
Report 8010047-2012-00159 · MDR key 2588945
Device
Generic name
Colonoscope
Manufacturer
Olympus Medical Systems CorporationModel number
CF-Q160ALCatalog number
CF-Q160AL
Lot number
NA
Product problems
- No Display/Image
Patient
UNK
- No Consequences Or Impact To Patient
Narrative
Description of Event or Problem
THE USER FACILITY REPORTED THAT DURING A DIAGNOSTIC COLONOSCOPY THE USERS EXPERIENCED A COMPLETE LOSS OF IMAGE. THE COLONOSCOPE WAS WITHDRAWN FROM THE PT AND THE PROCEDURE WAS COMPLETED USING A DIFFERENT BUT SIMILAR ENDOSCOPE. THERE WAS NO PT INJURY REPORTED.
Additional Manufacturer Narrative
THE DEVICE REFERENCED IN THIS REPORT WAS RETURNED TO OLYMPUS FOR EVAL. THE EVAL DID NOT CONFIRM THE USER'S REPORT OF IMAGE LOSS; HOWEVER, THERE WAS EVIDENCE OF FLUID INVASION IN THE ELECTRICAL CONNECTOR, SCOPE CONNECTOR, AND CONTROL BODY UNIT, WHICH LIKELY CAUSED THE IMAGE DIFFICULTY. THE DEVICE PASSED THE LEAK TEST. THE BENDING SECTION COVER AND BENDING SECTION COVER GLUE ARE FOUND TO BE A NON-OLYMPUS REPAIRS. THE OBJECTIVE LENS WAS CRACKED. THE SWITCH #1 BUTTON IS NOT WORKING. THE INSERTION TUBE FAILED THE INSULATION TEST. THE DEVICE WAS REFURBISHED. THIS REPORT IS BEING SUBMITTED AS A MEDICAL DEVICE REPORT IN AN EXCESS OF CAUTION.