inforMED
MalfunctionBSZ

CARESTATION 620

Received Jul 20, 2026 · Event occurred Jun 26, 2026

Report 9710602-2026-01823 · MDR key 25882913

Device

Generic name

Anesthesia Gas Machine

Model number

A2

Catalog number

NA

Lot number

NA

Product problems

  • Therapeutic or Diagnostic Output Failure

Patient

Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

IT WAS REPORTED THAT THERE WAS AN ERROR RESULTING IN LOSS OF MECHANICAL VENTILATION DURING PRE-USE CHECKOUT. THERE WAS NO PATIENT INVOLVEMENT.

Additional Manufacturer Narrative

A GE HEALTHCARE SERVICE REPRESENTATIVE PERFORMED A CHECKOUT OF THE SYSTEM AND CONFIRMED THE REPORTED ISSUE. THE ANESTHESIA CONTROL BOARD WAS REPLACED TO RESOLVE THE ISSUE.BLOCK A: NO REPORT OF PATIENT INVOLVEMENT. LEGAL MANUFACTURER:HCS MADISON - 3030 OHMEDA DR, USA MADISON, WI 53718.