MalfunctionBSZ
CARESTATION 620
Received Jul 20, 2026 · Event occurred Jun 26, 2026
Report 9710602-2026-01823 · MDR key 25882913
Device
Generic name
Anesthesia Gas Machine
Manufacturer
Ge Medical Systems (china) Co., LtdModel number
A2Catalog number
NA
Lot number
NA
Product problems
- Therapeutic or Diagnostic Output Failure
Patient
Unknown
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
IT WAS REPORTED THAT THERE WAS AN ERROR RESULTING IN LOSS OF MECHANICAL VENTILATION DURING PRE-USE CHECKOUT. THERE WAS NO PATIENT INVOLVEMENT.
Additional Manufacturer Narrative
A GE HEALTHCARE SERVICE REPRESENTATIVE PERFORMED A CHECKOUT OF THE SYSTEM AND CONFIRMED THE REPORTED ISSUE. THE ANESTHESIA CONTROL BOARD WAS REPLACED TO RESOLVE THE ISSUE.BLOCK A: NO REPORT OF PATIENT INVOLVEMENT. LEGAL MANUFACTURER:HCS MADISON - 3030 OHMEDA DR, USA MADISON, WI 53718.